The benchtop Optical Spectrum Analyzer MS9740B features wide dynamic range, high resolution, and fast sweep speeds over a wavelength range of 600 nm to 1750 nm. It supports multimode fiber input and is ideal for manufacturing and evaluating 850-nm band VCSEL modules.
As well as keeping the same functions and performance as its predecessor, the MS9740B slashes measurement processing times by up to half(Rx bandwidth setting: 1 kHz or 200 Hz) while overcoming problems of decreased measurement sensitivity caused by increased processing speed.
Moreover, it retains optical active device (LD-Module, DF-LD, FP-LD, LED, PMD, WDM, Opt Amp, and WDM Filter) measurement menu screens for evaluating devices. It supports all-at-once measurements of key evaluation items, such as optical center wavelength, level, OSNR, spectrum width, etc., and displays these results at one screen.
GPIB Interface, SMSR Measurement Time (DFB Light Source), VBW: 1 kHz_Fast (MS9740B)/1 kHz(MS9740A) Setting, 0.1-nm Resolution, 30-nm Sweep Width, 1001 Sample Points
Parameters / Characteristics | Specifications | |
---|---|---|
MS9740B | MS9740B-009 | |
Applicable optical fiber | SM fiber (ITU-T G.652), 50μm/125μm GI fiber | SM fiber (ITU-T G.652), 50μm/125μm GI fiber, 62.5μm/125μm GI fiber |
Optical connector | User replaceable: FC, SC, ST, DIN | |
Wavelength measurement range | 600 nm to 1750 nm | |
Wavelength accuracy*1 | ±20 pm (1520 nm to 1620 nm, Resolution: 0.03 to 0.2 nm)*2 ±100 pm (1520 nm to 1620 nm, Resolution: 0.5, 1.0 nm)*2 ±300 pm (600 nm to 1520 nm)*2 ±200 pm (1520 nm to 1570 nm)*2 ±300 pm (1570 nm to 1750 nm)*2 |
±50 pm (1530 nm to 1570 nm, Resolution: 0.07 to 0.2 nm)*2 ±100 pm (1530 nm to 1570 nm, Resolution: 0.5, 1.0 nm)*2 ±300 pm (1570 nm to 1750 nm)*4 |
Setting resolution | 0.03, 0.05, 0.07, 0.1, 0.2, 0.5, 1.0 nm (0.03 nm, 0.05 nm only 1550 nm band and room temperature) | 0.07, 0.1, 0.2, 0.5, 1.0 nm |
Maximum input power | +23 dBm (Optical ATT: On) | |
Dynamic Range*1 | High dynamic range: 70 dB (±1 nm from peak wavelength), 60 dB (±0.4 nm from peak wavelength), 42 dB (±0.2 nm from peak wavelength) Normal dynamic range: 62 dB (±1 nm from peak wavelength), 58 dB (±0.4 nm from peak wavelength), 42 dB (±0.2 nm from peak wavelength) [Wavelength: 1550 nm, Resolution: 0.05 nm, Optical Att: Off, 20° to 30°C] |
High dynamic range: 70 dB (±1 nm from peak wavelength, 20° to 30°C), 60 dB (±0.5 nm from peak wavelength, 20° to 30°C) 65 dB (±1 nm from peak wavelength, 5° to 45°C), 55 dB (±0.5 nm from peak wavelength, 5° to 45°C) Normal dynamic range: 62 dB (±1 nm from peak wavelength, 20° to 30°C), 58 dB (±0.5 nm from peak wavelength, 20° to 30°C) 57 dB (±1 nm from peak wavelength, 5° to 45°C), 53 dB (±0.5 nm from peak wavelength, 5° to 45°C) [Wavelength: 1550 nm, Resolution: 0.07 nm, using SM fiber, Optical Att: Off] |
Sweep*1 | Sweep time: ≤0.2 s (span: 5 nm, Resolution: 0.1 nm), ≤0.3 s (span: 500 nm) [VBW: 10 kHz, Normal dynamic range, center 1550 nm (span: 5 nm), 1200 nm (span: 500 nm), sweep start to stop, no optical input, sampling point: ≤501] Sweep time: 0.35 s / 30 nm (typ.) [VBW: 1 kHz- Fast, dynamic range, center 1550 nm, Resolution: 0.1 nm, sweep start to stop, optical input -10 dBm, sampling point: 1001] Sweep time: 1,65 s / 30 nm (typ.) [VBW: 200 Hz-Fast, dynamic range, center 1550 nm, Resolution: 0.1 nm, sweep start to stop, optical input -10 dBm, sampling point: 1001] |
|
Dimensions and Mass | 426 (W) × 177 (H) × 350 (D) mm (excluding projections) ≤15.0 kg (without option) |
*1: Warm-up the instrument for at least 2 hours before measurement by performing repeated sweeping at span ≥ 100 nm, VBW ≥ 10 kHz. Perform waveform calibration after auto-optical alignment (WI Cal) and keep the instrument at the same temperature unless stated otherwise. Use either SM fiber (ITU-T G.652) or GI fiber (50μm/125μm) with a return loss of >40 dB, or GI fiber (62.5μm/125μm) with a return loss of >38 dB
*2: Built-in MS9740B-002, after Wl cal (ref) wavelength calibration execution, at stable room temperature
*3: After Wl cal (Ext) wavelength calibration execution by external light source, such as Single Longitudinal mode laser (DFB-LD), using SM fiber
*4: After Wl cal (Ext) wavelength calibration execution by external light source, such as DFB-LD, using SM fiber or GI fiber (50μm/125μm or 62.5μm/125μm)
Checking with Anritsu...
Model Number | Setting Resolution | Wavelength measurement range | Maximum input power |
---|---|---|---|
![]() |
0.03, 0.05, 0.07, 0.1, 0.2, 0.5, 1.0 nm (0.03 nm, 0.05 nm only 1550 nm band and room temperature) | 600 nm to 1750 nm | +23 dBm (Optical ATT: On) |
Note: All trademarks that appear on this website are the property of their respective owners.